Image Intensifier Tube || IITube|| Fluoroscopy
Summary
TLDRThe Image Intensifier Tube (IIT) is a key component in fluoroscopy systems, converting low-intensity X-rays into high-intensity light images. The process begins when X-rays pass through the patient and are absorbed by the input phosphor, which emits light. This light interacts with a photocathode to produce photoelectrons, which are then accelerated towards the output phosphor. The output phosphor generates a much brighter light image, which is captured for medical analysis. The IIT creates an inverted image due to the focal point of the electron flow, enhancing the clarity and brightness of fluoroscopic images.
Takeaways
- 😀 The Image Intensifier Tube (IIT) is an electronic device used in fluoroscopy systems to enhance x-ray images.
- 😀 The IIT is a vacuum tube placed inside a metal housing that converts low-intensity x-rays into high-intensity light.
- 😀 The x-ray beam transmitted through the patient enters the IIT through an input window, where it is absorbed by the input phosphor.
- 😀 The input phosphor is a fluorescent screen that converts x-ray photons into light photons (visible light).
- 😀 A photocathode, coated on the inner surface of the input phosphor, emits photoelectrons when struck by light.
- 😀 The number of photoelectrons emitted by the photocathode is proportional to the brightness of the input phosphor.
- 😀 The anode (or accelerating anode) is a metal disk that accelerates the emitted photoelectrons towards the output phosphor.
- 😀 Electrostatic focusing lenses focus the electron beam toward the output phosphor, allowing for a clearer image.
- 😀 The accelerated electrons strike the output phosphor, producing a much brighter light image compared to the input phosphor.
- 😀 Due to all electrons passing through a common focal point, the image generated on the output phosphor is inverted.
- 😀 The bright light produced by the output phosphor is captured by image-capturing devices like vidicon cameras for analysis.
Q & A
What is an Image Intensifier Tube (IIT)?
-An Image Intensifier Tube (IIT) is an electronic device used in fluoroscopy systems to convert low-intensity, attenuated X-rays into high-intensity light images.
How does the Image Intensifier Tube work?
-The IIT functions by converting the energy of X-rays into visible light using a series of components: the X-ray photons enter the IIT through the input window and are absorbed by the input phosphor, which then emits light photons. These light photons are then converted into electrons and accelerated toward the output phosphor, where they produce a high-intensity light image.
What is the role of the input phosphor in an Image Intensifier Tube?
-The input phosphor is a fluorescent screen that absorbs the transmitted X-ray photons and emits light photons, effectively converting the energy of the X-rays into visible light.
What is a photocathode, and how does it function in the Image Intensifier Tube?
-The photocathode is a thin metal layer coated on the inner surface of the input phosphor. When light from the input phosphor strikes the photocathode, it emits photoelectrons in numbers proportional to the brightness of the light.
What is the purpose of the anode in the Image Intensifier Tube?
-The anode, also known as the accelerating anode, is a circular metal piece with a hole in the center. Its role is to accelerate the electrons emitted by the photocathode toward the output phosphor, contributing to the creation of the final high-intensity image.
How do electrostatic focusing lenses work in the Image Intensifier Tube?
-Electrostatic focusing lenses are negatively charged electrodes that focus the electron beam toward the output phosphor, ensuring that the accelerated electrons hit the phosphor in a focused manner.
What happens when the accelerated electrons hit the output phosphor?
-When the highly accelerated electrons hit the output phosphor, they produce a large number of light photons, resulting in a much brighter image compared to the input phosphor.
What is the role of the output phosphor in the Image Intensifier Tube?
-The output phosphor is where the accelerated electrons from the photocathode are converted into a high-intensity light image. This image is then captured by an image capturing device, such as a vidicon camera.
Why does the Image Intensifier Tube create an inverted image?
-The Image Intensifier Tube creates an inverted image because all the electrons pass through a common focal point, which leads to the inversion of the image on the output phosphor.
What is the significance of using an Image Intensifier Tube in fluoroscopy?
-The IIT is crucial in fluoroscopy because it enhances the brightness of the image, allowing clearer visualization of internal structures, even in low-light conditions, by converting low-intensity X-rays into visible light.
Outlines
This section is available to paid users only. Please upgrade to access this part.
Upgrade NowMindmap
This section is available to paid users only. Please upgrade to access this part.
Upgrade NowKeywords
This section is available to paid users only. Please upgrade to access this part.
Upgrade NowHighlights
This section is available to paid users only. Please upgrade to access this part.
Upgrade NowTranscripts
This section is available to paid users only. Please upgrade to access this part.
Upgrade NowBrowse More Related Video
The Image Intensifier Tube
RADT 086 Image Intensification Tube
Gamma camera | Components & Function l Visual explanation
Biomedical Instrumentation- X-RAY Machine
What are Real and Virtual Images? | Reflection of Light | Infinity Learn
CT scan | computerized tomography (CT) scan |What is a CT scan used for? | Clinical application
5.0 / 5 (0 votes)