Medonic Hematology M32 Maintenance
Summary
TLDRThis tutorial provides a step-by-step guide for maintaining the Medonic M32 analyzer, covering four key maintenance intervals: daily, weekly, monthly, and 6-monthly. Each procedure ensures optimal performance by using specialized reagents like Diluent, Lyze, and Hypochlorite cleaner. The daily maintenance focuses on flushing the system and cleaning intake pipes, while the weekly cleaning targets residue removal. Monthly maintenance includes enzymatic cleaning, and the 6-month maintenance utilizes a comprehensive cleaning kit to ensure the instrument is free of contaminants. These procedures are essential for keeping the M32 analyzer in top condition for accurate measurements.
Takeaways
- 😀 Daily maintenance includes flushing the system using the 'Prime System' option to ensure no debris or blood residues are left in the tubing.
- 😀 Always clean the intake needles with a tissue soaked in 70% ethanol at the start and end of each day to maintain good condition.
- 😀 Weekly maintenance involves using 2% Hypochlorite Cleaner to clean the system, ensuring no protein or blood residues remain in the tubing.
- 😀 The weekly cleaning should use a special 2% Hypochlorite solution, not commercial household cleaners, to avoid damaging components.
- 😀 For monthly maintenance, use the Enzymatic Cleaner for 15 minutes to digest blood residues, followed by background flushing to remove any leftover debris.
- 😀 During monthly cleaning, clean the Autoloader by placing a sealed tube with the cleaner into the sample position and selecting 'Clot Prevention.'
- 😀 The 6-monthly maintenance involves a thorough cleaning process with three types of cleaners: Enzymatic Cleaner, 2% Hypochlorite solution, and Detergent Cleaner.
- 😀 Empty the system before beginning the 6-monthly maintenance procedure to ensure all liquids are removed from the instrument.
- 😀 Always clean the Diluent and Lyze probes carefully with tissue to avoid contamination from previous reagents before reusing them.
- 😀 After 6-monthly maintenance, verify that the system is fully functional by completing the 'Fill System' and 'Background' flushing procedures.
Q & A
What is the primary purpose of the Daily Maintenance procedure for the Medonic M32 analyzer?
-The primary purpose of the Daily Maintenance is to ensure that the system is free from debris, salt, and blood residues, ensuring accurate measurements for subsequent tests.
How long does the 'Prime System' procedure take, and what does it do?
-The 'Prime System' procedure takes about 1 minute and flushes all tubing and over 20 valves in the system to prepare it for measurements.
What should the blank values (e.g., WBC, RBC, HGB, PLT) show after performing the Daily Maintenance?
-After performing the Daily Maintenance, all blank values such as WBC, RBC, HGB, and PLT should show '0' to ensure the system is clean and ready for accurate measurements.
Why is it important to clean the inlet sipper pipe with 70% ethanol daily?
-Cleaning the inlet sipper pipe with 70% ethanol daily ensures that no blood residues or debris remain on the intake needles, which could affect the accuracy of the measurements.
What is the purpose of using the 2% Hypochlorite Cleaner in the Weekly Maintenance?
-The 2% Hypochlorite Cleaner is used in the Weekly Maintenance to remove blood residues, particularly proteins, from the system, ensuring it stays clean and functional.
Can household hypochlorite cleaners be used for the Weekly Maintenance?
-No, commercial household hypochlorite cleaners should not be used because they contain perfumes and colorants that can damage the tubing and valves in the instrument.
What does the 'Clot Prevention' procedure during Monthly Maintenance do?
-The 'Clot Prevention' procedure helps digest and remove blood residues from the Autoloader's intake system, ensuring it remains clean and free of blockages.
How long does the 'Clot Prevention' procedure take in the Monthly Maintenance?
-The 'Clot Prevention' procedure takes about 15 minutes to complete.
What are the three cleaners included in the Boule Cleaning Kit for 6-Monthly Maintenance?
-The three cleaners in the Boule Cleaning Kit for the 6-Monthly Maintenance are Enzymatic Cleaner, 2% Hypochlorite solution, and Detergent Cleaner.
What should be done after the 6-Monthly Maintenance procedure is completed?
-After the 6-Monthly Maintenance procedure is completed, the system should be filled with Diluent and Lyze reagents, and a BACKGROUND flush should be performed to verify the instrument is ready for use.
Outlines
Этот раздел доступен только подписчикам платных тарифов. Пожалуйста, перейдите на платный тариф для доступа.
Перейти на платный тарифMindmap
Этот раздел доступен только подписчикам платных тарифов. Пожалуйста, перейдите на платный тариф для доступа.
Перейти на платный тарифKeywords
Этот раздел доступен только подписчикам платных тарифов. Пожалуйста, перейдите на платный тариф для доступа.
Перейти на платный тарифHighlights
Этот раздел доступен только подписчикам платных тарифов. Пожалуйста, перейдите на платный тариф для доступа.
Перейти на платный тарифTranscripts
Этот раздел доступен только подписчикам платных тарифов. Пожалуйста, перейдите на платный тариф для доступа.
Перейти на платный тарифПосмотреть больше похожих видео
5.0 / 5 (0 votes)