Training Video MEDUMAT Easy CPR | WEINMANN Emergency

WEINMANN Emergency
6 Apr 202012:53

Summary

TLDRCarson Luo Tian introduces Weinman Emergency's transport ventilator, the 'Maquet easy CPR', designed for emergency and transport situations. The ventilator is automated, with a supplemental inhalation function, capable of administering controlled ventilation for patients weighing 10kg and above. It features a user-friendly interface with color-coded settings for different patient groups and offers modes for mask ventilation, CPR, and demand flow oxygen supply. The device is mounted on a life base-3 system for safe transport and includes a 2-liter oxygen cylinder. The script also covers operation, maintenance, and functional checks for the device.

Takeaways

  • 🌟 The speaker introduces the emergency and transport ventilator, named 'Muhammad easy CPR', which is mounted on a carrying system called 'life base-3'.
  • πŸ₯ Muhammad easy CPR is an automated, of-seat ventilator with a supplemental inhalation function, suitable for patients weighing 10 kilograms and more.
  • πŸš‘ The device can administer controlled ventilation, trigger a single ventilator breath during CPR, and deliver breath-managed oxygen inhalation with 100% oxygen in demand flow mode.
  • 🎈 Life base-3 can accommodate and safely transport an oxygen cylinder containing 400 liters of oxygen, with a zippered access for all necessary components.
  • πŸ”Œ The major components of the portable system include the ventilator, patient whole system, ventilation mask, middle trigger, 2-liter oxygen cylinder, pressure reducer, and oxygen quick coupling.
  • πŸ”‘ The device is operated via an on-off key and a regulator knob, with color coding indicating settings suitable for adult, child, or infant patient groups.
  • πŸ›‘ The green LED lamp indicates the current mode of use, and the LED ventilation pressure gauge shows the current ventilation pressure.
  • 🚨 Alarm indicators include flashing red LED signals and audio responses for various issues such as exceeding ventilation pressure, disconnection from patient oxygen supply, or insufficient pressure at the device inlet.
  • πŸ”‹ Muhammad easy CPR is equipped with a 3.6-volt main battery that needs replacement every two years or earlier if necessary, using batteries from Weinman emergency.
  • πŸ”„ The device can be used in demand flow mode to administer 100% oxygen to a spontaneously breathing patient, with a simple switch on and mask application.
  • 🧼 Regular maintenance and cleaning are recommended, including wiping components with disinfectant, cleaning the protective bag in a washing machine, and conducting a functional check every six months or more frequently.

Q & A

  • What is the name of the product specialist presenting the ventilator?

    -The product specialist's name is Carson Luo Tian.

  • What is the full name of the ventilator being presented?

    -The ventilator being presented is called 'Muhammad easy CPR'.

  • What is the primary function of the Muhammad easy CPR ventilator?

    -The primary function of the Muhammad easy CPR ventilator is to administer controlled ventilation to patients weighing 10 kilograms and more, trigger a single ventilator breath during cardiopulmonary resuscitation (CPR), and deliver breath-managed oxygen inhalation with 100% oxygen in demand flow mode.

  • What is the maximum weight capacity for patients using the Muhammad easy CPR ventilator?

    -The Muhammad easy CPR ventilator can be used for patients weighing 10 kilograms and more.

  • What is the purpose of the life base-3 carrying system?

    -The life base-3 carrying system is designed to accommodate and safely transport an oxygen cylinder containing 400 liters of oxygen.

  • How is the life base-3 carrying system opened for access to components?

    -The life base-3 carrying system can be opened with a zipper for access to all necessary components.

  • What color coding is used for the ventilator parameters and what do they represent?

    -The color coding for the ventilator parameters includes brown for adult patient group, orange for child patient group, and yellow for infant patient group.

  • What does the green LED lamp indicate on the Muhammad easy CPR ventilator?

    -The green LED lamp indicates the current mode of use on the Muhammad easy CPR ventilator.

  • How can you switch off the acoustic alarm on the Muhammad easy CPR ventilator?

    -You can switch off the acoustic alarm by pressing briefly the alarm mute key.

  • What is the purpose of the CPR key on the Muhammad easy CPR ventilator?

    -The CPR key is used to end automatic ventilation and to continue manual ventilation with the Muhammad easy CPR ventilator.

  • How often should the main battery of the Muhammad easy CPR ventilator be replaced?

    -The main battery of the Muhammad easy CPR ventilator should be replaced every two years as a part of preventative maintenance.

  • What is the recommended procedure for cleaning the Muhammad easy CPR ventilator?

    -The Muhammad easy CPR ventilator should be cleaned by wiping the middle trigger and pressure reducer with a disinfectant, and the protective bag may be cleaned in the washing machine at 30 degrees Celsius as needed. After disinfecting, rinse all parts with clear cold water and let them air dry.

Outlines

00:00

πŸš‘ Introduction to the Emergency Ventilator

Carson Luo Tian, a product specialist at Weinman Emergency, introduces the emergency and transport ventilator, MiniMott Easy CPR, mounted on a carrying system called Life Base-3. This automated ventilator is capable of controlled ventilation for patients weighing 10kg and more, single ventilator breaths during CPR, and demand flow mode for 100% oxygen inhalation. Life Base-3 includes a ventilator, patient whole system, ventilation mask, middle trigger, 2-liter oxygen cylinder, pressure reducer, and a T-shaped distributor with oxygen quick coupling. The device is operated using an on-off key and a regulator knob with color coding for different patient groups. It also features an LED ventilation pressure gauge and alarm indicators for various issues such as ventilation pressure, disconnection, and low pressure or battery.

05:04

πŸ”‹ Battery and Maintenance Information

The MiniMott Easy CPR is powered by a 3.6-volt main battery, which should be replaced every two years for preventative maintenance. If replacement is needed earlier, only Weinman Emergency batteries should be used. The device emits audio responses and alarm signals through a loudspeaker. The speaker and middle trigger plug are detailed, including how to attach and remove them without damaging the socket. Instructions are also provided for detaching the pressure gauge tube and connecting the ventilation hole. The device can be used in demand flow mode to administer 100% oxygen to spontaneously breathing patients. The device always starts in mask ventilation mode, and CPR mode can be activated by pressing the CPR button and attaching a ventilation mask. For intubated patients, guidelines recommend uninterrupted chest compressions with continuous ventilation at a fixed frequency.

10:05

🧼 Cleaning and Functional Check Procedures

The MiniMott Easy CPR portable system, including the Life Base 3 oxygen cylinder, middle trigger, and pressure reducer, can be cleaned by wiping with a disinfectant. The protective bag is machine washable at 30 degrees Celsius. The patient host system should be cleaned with a brush and disinfectant solution, followed by rinsing with clear cold water and air drying. Defective or damaged parts should be replaced. A functional check must be performed before each use, after each dismantling and reassembly, and at least every six months as specified in the instructions. If the device fails the functional check, it should be tested by a Weinmann emergency authorized service technician.

Mindmap

Keywords

πŸ’‘Ventilator

A ventilator is a medical device that assists in the process of respiration by delivering breaths to a patient who cannot breathe or is breathing insufficiently. In the video, the 'Maquet easy CPR' is a specific type of ventilator presented, designed for emergency and transport situations, emphasizing its automated seating ventilator with supplemental inhalation function.

πŸ’‘Cardiopulmonary Resuscitation (CPR)

CPR is an emergency procedure performed when the heart has stopped beating, and the person is not breathing. The script mentions that the ventilator can be used during CPR to administer a single ventilator breath, which is crucial for maintaining oxygenation during resuscitation efforts.

πŸ’‘Life Base-3

Life Base-3 is the carrying system for the Maquet easy CPR ventilator, designed to accommodate and safely transport an oxygen cylinder. It is an integral part of the portable system, allowing for easy access to all necessary components and facilitating transport during emergencies.

πŸ’‘Oxygen Cylinder

An oxygen cylinder is a container for storing and transporting oxygen under pressure. The script specifies that Life Base-3 can hold an oxygen cylinder containing 400 liters of oxygen, which is essential for the ventilator's operation, especially in providing 100% oxygen during demand flow mode.

πŸ’‘Pressure Reducer

A pressure reducer is a device that regulates the pressure of oxygen from the cylinder to a safe and usable level for the ventilator. The script mentions a T-shaped distributor and oxygen quick coupling type, which connect to the ventilator to ensure the correct pressure for oxygen delivery.

πŸ’‘Ventilation Mask

A ventilation mask is used to deliver oxygen or anesthetic gases to a patient during respiratory support or general anesthesia. In the context of the video, the mask is part of the patient whole system for the Maquet easy CPR ventilator, used for administering ventilation to patients.

πŸ’‘Regulator Knob

The regulator knob is used to adjust the ventilator parameters, such as the ventilation pressure and frequency. The script describes how to set these parameters using the color-coded markings on the knob, which helps in selecting the appropriate settings for adult, child, or infant patients.

πŸ’‘LED Indicator

LED indicators are used in medical devices to provide visual feedback on the device's status and settings. The script mentions a green LED lamp that indicates the current mode of use, and an LED ventilation pressure gauge that displays the current ventilation pressure.

πŸ’‘Alarm System

An alarm system in medical devices alerts the user to potential issues or malfunctions. The script describes different alarm indicators, such as flashing red LEDs and audio responses, which inform the user of issues like exceeded ventilation pressure, disconnection from the patient oxygen supply, or insufficient pressure at the device inlet.

πŸ’‘CPR Mode

CPR mode is a specific setting on the ventilator that allows for the delivery of breaths in conjunction with chest compressions during CPR. The script explains how to activate CPR mode by pressing the CPR key and attaching a ventilation mask to the patient, enabling manual ventilation with the automatic ventilator.

πŸ’‘Demand Flow Mode

Demand flow mode is a setting on the ventilator that allows a patient to receive 100% oxygen spontaneously. The script describes how to administer oxygen in this mode by switching on the device and adjusting the regulator knob, which is essential for patients who are breathing on their own but require supplemental oxygen.

πŸ’‘Functional Check

A functional check is a routine inspection performed to ensure that the ventilator is working correctly. The script emphasizes the importance of conducting a functional check before each use, after each dismantling and reassembly, and at least every six months, to maintain the device's reliability and safety.

Highlights

Introduction of the emergency and transport ventilator, Muhammad Easy CPR.

The ventilator is mounted on a carrying system, Life Base-3, for easy transportation.

Muhammad Easy CPR is an automated, of seating ventilator with in, supplemental inhalation function.

It can administer controlled ventilation to patients weighing 10 kilograms and more.

Trigger a single ventilator breath during cardiopulmonary resuscitation (CPR).

Deliver breath-managed oxygen inhalation with 100% oxygen in demand flow mode.

Life Base-3 can accommodate an oxygen cylinder containing 400 liters of oxygen.

The system can be opened with a zipper for access to all necessary components.

Major components include the ventilator, patient whole system, ventilation mask, and middle trigger.

The 2-liter oxygen cylinder and pressure reducer are part of the system.

Quick coupling type Volta to connect to an external gas supply.

Device is switched on and off with the on-off key and ventilator parameters are set with a regulator knob.

Color coding indicates suitable settings for adult, child, and infant patient groups.

Green LED lamp indicates the current mode of use.

Alarm indicator provides visual and audio responses for various alarm conditions.

Acoustic alarm can be muted, and CPR mode can be activated with a dedicated key.

3.6-volt main battery requires replacement every two years or as needed.

Loudspeaker emits audio responses and alarm signals.

Instructions for attaching and removing the middle trigger plug.

How to release and properly lock the portable system into the wall mounting base.

Use of demand flow mode to administer 100% oxygen to spontaneously breathing patients.

Initial operation mode is mask ventilation with pre-selected ventilation frequency and tidal volume.

CPR mode activation for manual ventilation using an automatic ventilator.

Guidelines for uninterrupted chest compressions and continuous ventilation during CPR.

Cleaning and maintenance instructions for the Muhammad Easy CPR system.

Functional check must be performed before each use, after reassembly, and at least every six months.

If the device fails the functional check, it should be tested by an authorized service technician.

Transcripts

play00:07

[Music]

play00:19

[Music]

play00:26

hello my name is Carson Luo Tian's I'm a

play00:28

product specialist at Wineman emergency

play00:31

here in Hamburg today I would like to

play00:33

present to you our emergency and

play00:36

transport ventilator Muhammad easy CPR

play00:38

mounted on his carrying system life

play00:41

base-3 Mohammed easy CPR is an automated

play00:45

of seating ventilator with in

play00:47

supplemental inhalation function with

play00:50

minimal easy CPR you can do the

play00:52

following administer controlled

play00:55

ventilation to patients weighing 10

play00:58

kilogram and more trigger a single

play01:02

ventilator breath during cardiopulmonary

play01:04

resuscitation CPR and deliver breath

play01:08

managed oxygen inhalation with 100%

play01:11

oxygen in demand flow mode life based 3

play01:16

accommodates and safely transports an

play01:18

oxygen cylinder containing 400 liters of

play01:21

oxygen

play01:30

life pastry can be opened with a zipper

play01:34

for access to all necessary components

play01:43

the major components of the portable

play01:46

system life base-3 are the ventilator

play01:50

minimott easy CPR with patient whole

play01:55

system

play02:00

ventilation mask and middle trigger

play02:12

the 2-liter oxygen cylinder

play02:21

the pressure reducer obviously very fast

play02:23

to with t-shaped distributor and oxygen

play02:27

quick coupling type Volta to connect to

play02:30

an external gas supply the pressure hose

play02:37

between the pressure reducer of the way

play02:39

fast too and the ventilator Mohammed

play02:42

easy CPR

play02:50

you

play02:52

the device is switched on and off with

play02:55

the on-off key to set the ventilator

play02:58

parameters turn the regulator knob with

play03:02

the markings to the required position

play03:05

the color coding shows you whether the

play03:08

settings on the knob are suitable for

play03:10

your patient Brown adult patient group

play03:15

orange child patient group yellow infant

play03:22

patient group the switch for mask cube

play03:27

ventilation regulates the maximum

play03:29

ventilation pressure for the selected

play03:32

setting the maximum ventilation pressure

play03:35

is 20 milli bar for mask ventilation and

play03:38

45 millibar for tube ventilation the

play03:43

green LED lamp tells you which mode is

play03:46

currently in use you can read the

play03:48

current ventilation pressure from the

play03:51

LED ventilation pressure gage

play03:55

the alarm indicator is divided into

play03:58

separate alarm areas stenosis flashing

play04:03

red LED alarm signal and audio response

play04:06

tell you if the previous reset

play04:08

ventilation pressure has been exceeded

play04:13

disconnection flashing red LED alarm

play04:18

signal and audio response tell you when

play04:20

Mohammed ICCPR

play04:21

disconnects from patient oxygen supply

play04:27

pressure below 2.7 bar flashing red LED

play04:33

alarm signal and audio response indicate

play04:36

that there is insufficient pressure at

play04:39

the device inlet battery flashing red

play04:43

LED signals insufficient battery charge

play04:47

you can switch off the acoustic alarm by

play04:50

pressing briefly the alarm mute key

play04:53

press the CPR key to end the automatic

play04:57

ventilation and to continue manual

play04:59

ventilation with me to trigger the green

play05:03

LED shows you that you are working in

play05:05

CPR mode

play05:10

this is where the pressure hose is

play05:12

attached miramar easy CPR is equipped

play05:16

with a 3.6 volt main battery this

play05:21

battery is replaced every two years as a

play05:24

part of preventative maintenance if

play05:27

however the battery has to be replaced

play05:30

earlier then use only a battery that

play05:32

comes from Weinman emergency the

play05:36

loudspeaker emits audio responses and

play05:40

alarm signals attach the middle trigger

play05:49

plug here

play06:00

to remove pull the plaque downwards make

play06:06

sure you do not turn the plaque while

play06:08

removing it because you could damage the

play06:10

socket

play06:15

slide the pressure gage tube onto its

play06:18

connection then connect the ventilation

play06:23

hole

play06:29

you

play06:35

to release live pastry from the wall

play06:37

mounting base station

play06:39

grasp the handle firmly with one hand

play06:42

and pulled the release lever upwards tip

play06:48

the portable system slightly toward the

play06:51

front and remove the system when you put

play06:58

the life pastry back in the base station

play07:00

make sure it is properly locked that

play07:03

means that the red knob on the locking

play07:05

mechanism should not extend beyond the

play07:07

housing

play07:15

you

play07:19

you can use minimott easy CPR in demand

play07:22

flow mode to administer 100% oxygen to a

play07:26

patient who was spontaneously breathing

play07:30

switch on the device and turn the

play07:37

regulator knob clockwise until it passes

play07:40

the point of resistance and the green

play07:43

LED lights up

play07:48

hold the mask over the patient's nose

play07:50

and mouth and tell the patient to

play07:52

continue breathing calmly and evenly

play08:02

you

play08:05

you

play08:08

if you switch on minimott easy CPR it

play08:11

always starts in mask ventilation mode

play08:14

using your pre-selected ventilation

play08:16

frequency and tidal volume

play08:21

you

play08:33

you

play08:42

if you need to perform a cardiopulmonary

play08:45

resuscitation CPR

play08:47

you should run Miramar easy CPR and in

play08:50

CPR mode to activate CPR mode simply

play08:56

press the button CPR and attach a

play09:01

ventilation mask to the patient house

play09:05

now you're ready to perform manual

play09:08

ventilation by using an automatic

play09:11

ventilator

play09:17

[Music]

play09:26

[Music]

play09:53

[Music]

play10:02

if you have intubated the patient the

play10:05

guidelines on resuscitation recommend

play10:07

uninterrupted chest compressions

play10:10

accompanied at the same time by

play10:12

continuous ventilation with a fixed

play10:14

frequency in this case please deactivate

play10:17

the manual ventilation by pushing the

play10:20

CPR button and changing the mask tube

play10:23

selector over to tube mental ation the

play10:26

green LED next to the laryngoscope

play10:29

symbol shows the correct switch over

play10:40

can keep clean the middle mud easy CPR

play10:43

portable system life base 3 oxygen

play10:47

cylinder me to trigger and pressure

play10:50

reducer by wiping them with a

play10:53

disinfectant

play10:56

the protective bag may be cleaned and

play10:58

the washing machine at 30 degrees

play11:00

Celsius maximum as needed remove a

play11:06

visible dirt from the patient host

play11:09

system with a brush and immerse it in a

play11:11

disinfectant solution observe the

play11:14

soaking time important after the

play11:19

disinfecting procedure rinse all parts

play11:22

with clear cold water and let them air

play11:25

dry if you see parts that are defective

play11:31

or damaged replace them

play11:41

you

play11:43

before each use after each dismantling

play11:46

and reassembly and at the very least

play11:48

every six month the user must carry out

play11:51

a functional check on the ventilator

play11:54

make the functional check as specified

play11:57

in the instructions for use if the

play12:00

device does not pass the functional

play12:02

check repeated several times if the

play12:06

device still does not pass have the

play12:09

device tested by a service technician

play12:11

authorized by Weinmann emergency

play12:21

you

play12:49

[Music]

Rate This
β˜…
β˜…
β˜…
β˜…
β˜…

5.0 / 5 (0 votes)