Analisis kuantitatif potensiometri
Summary
TLDRThis video explains two potentiometric methods for quantitative analysis: the Calibration Curve Method and the Standard Addition Method. The first method involves creating a calibration curve by measuring the potential of standard solutions with varying concentrations of an analyte (e.g., calcium ions) and using regression equations to calculate unknown concentrations. The second method, Standard Addition, involves adding a known quantity of a standard solution to the sample, measuring the change in potential, and calculating the analyte concentration based on this change. Both methods emphasize the importance of accurate potential measurement and sample consistency.
Takeaways
- π The calibration curve method is used to determine the concentration of ions in a sample by measuring the cell potential with a selective electrode.
- π To apply the calibration curve method, prepare standard solutions of varying known concentrations and measure the potential using a selective electrode.
- π A calibration curve is created by plotting the measured potential against the concentration, allowing for the determination of unknown concentrations using regression analysis.
- π In the calibration curve method, it is crucial to ensure that the ionic strength of the sample and standard solutions is the same to avoid errors.
- π The second method is the standard addition method, which involves adding a known quantity of standard solution to a sample and measuring the change in potential.
- π The standard addition method uses the equation for the potential of the selective electrode to calculate the concentration of the analyte in the sample.
- π When using the standard addition method, subtract the initial potential from the potential after adding the standard solution to calculate the concentration of the analyte.
- π The regression equation from the calibration curve can be used to calculate the concentration of calcium ions (or any other ions) from the measured potential.
- π The standard addition method is especially useful when the sample matrix might interfere with the measurement, as it accounts for the background concentration of the analyte.
- π Both methods (calibration curve and standard addition) are reliable ways to perform quantitative analysis potentiometrically, with each suited to different types of samples and conditions.
Q & A
What are the two main methods used in potentiometric quantitative analysis?
-The two main methods are the calibration curve method and the standard addition method.
How is the calibration curve method used to determine the concentration of calcium ions?
-In the calibration curve method, a series of standard solutions with known concentrations of calcium ions is prepared. The potential is measured using a calcium ion-selective electrode, and the data is plotted with concentration on the x-axis and potential on the y-axis. A regression equation is generated, allowing the concentration of calcium in an unknown sample to be calculated.
Why is it important to maintain the same ionic strength in all solutions during the calibration curve method?
-Maintaining the same ionic strength in all solutions is crucial because variations in ionic strength can lead to errors in the potentiometric measurements, affecting the accuracy of the concentration determination.
What is the role of the calcium ion-selective electrode in the calibration curve method?
-The calcium ion-selective electrode measures the cell potential (voltage) for each standard solution. The potential varies with the concentration of calcium ions, and this relationship is used to create a calibration curve for determining the unknown concentration.
How does the standard addition method work?
-In the standard addition method, a known volume of standard solution with a known concentration of the analyte (e.g., nitrate) is added to a sample. The potential is measured before and after the addition, and the change in potential is used to calculate the concentration of the analyte in the sample.
What type of electrode is used in the standard addition method for nitrate determination?
-A nitrate ion-selective electrode is used in the standard addition method to measure the potential of the sample solution before and after adding the standard solution.
How is the concentration of nitrate determined in the standard addition method?
-The concentration of nitrate in the sample is determined by measuring the change in potential before and after adding a known volume of a standard nitrate solution. The Nernst equation is used to relate the potential to the concentration, allowing for the calculation of the nitrate concentration in the sample.
Why is the standard addition method preferred when sample matrices might interfere with measurements?
-The standard addition method is preferred in such cases because it compensates for interference from the sample matrix. By adding a known amount of standard solution, the impact of the matrix can be effectively accounted for in the calculation.
What equation is used in the standard addition method to calculate the analyte concentration?
-The equation used is based on the Nernst equation: Potential = k Γ charge Γ log(concentration). This equation is applied to calculate the concentration of the analyte based on the change in potential after the standard addition.
What is the significance of the change in potential in the standard addition method?
-The change in potential indicates the amount of analyte added to the sample. By measuring this change, the concentration of the analyte in the original sample can be calculated using the known volume of added standard solution and its concentration.
Outlines

This section is available to paid users only. Please upgrade to access this part.
Upgrade NowMindmap

This section is available to paid users only. Please upgrade to access this part.
Upgrade NowKeywords

This section is available to paid users only. Please upgrade to access this part.
Upgrade NowHighlights

This section is available to paid users only. Please upgrade to access this part.
Upgrade NowTranscripts

This section is available to paid users only. Please upgrade to access this part.
Upgrade NowBrowse More Related Video

Part 1 : Analisa Kualitatif dan Kuantitatif pada HPLC

Total Flavonoid Levels Using UV-Vis Spectrophotometer

Penelitian Mixed Methods (Penelitian Campuran): Kapan Menggunakannya? | Tirta Mursitama

Analisis Kualitatif & Kuantitatif Bahan Baku & Tablet Alopurinol secara Spektrofotometri UV Visible

SPEKTROFOTOMETER UV-VIS THERMO-GENESYS150 | FARMASI ANALISIS - STFI Bandung

Analisis Farmasi II: Analisa Bahan Aktif dalam Sediaan Semisolid
5.0 / 5 (0 votes)